QualiGuardQuality & Food Safety Consultancy

Part 2 of 6 · ISO/IEC 17025

Laboratory gap analysis: turn a checklist into an action plan

Gather evidence along the sample journey, prioritize gaps and define meaningful action closure.

QualiGuard Editorial Team2 min read
Laboratory pipette, microscope and test tubes

Examine practice alongside documents

Ask whether staff understand the work, whether practice matches procedures and whether records support the decisions made. General assurances in meetings are a weak substitute for specific examples.

As a suggested approach, follow one sample through request review, receipt, preparation, measurement, result review and reporting. Expand the assessment to the wider project scope.

Make observations actionable

‘Equipment management is weak’ gives an action owner little direction. ‘No record was found of the calibration certificate review for the balance used in the selected test’ identifies the evidence examined and what is missing.

Distinguish missing evidence from proof that an activity was not performed. Seek relevant records before concluding, and check requirement references against the authorized standard and applicable body policies.

Prioritize by impact

We suggest separating issues that may directly affect results, issues that weaken detection of problems, and organizational improvements. This is a project planning tool, not an official nonconformity grading system.

Where previously reported results may be affected, use the laboratory's procedures to evaluate the impact and decide appropriate action. Revising a form for future work does not address past effects.

  • Observation and supporting evidence.
  • Potential impact and affected activities.
  • Owner and target date.
  • Agreed action and effectiveness check.

Close with evidence

A revised procedure alone does not show successful implementation. Agree a suitable check: observing work, examining new records or reviewing a practical exercise.

For a certificate review gap, a suggested check is to examine the responsible person's application of the new approach to a later certificate. Record the decision, rationale and any further follow-up.

Sources and further reading

These articles offer educational guidance and suggested practical examples. They do not replace the standard or your accreditation body's requirements and do not confer accreditation.

Bring the learning into your laboratory

Discuss a training and qualification programme suited to your laboratory's activities and your team's experience.

Explore laboratory training and qualification